Download Agilent Technologies 4263B User`s guide
Transcript
4263B 5. Set the test signal level to 100 mV. Press . , and press Set the level to 100 mV using the numeric keys or 6. Perform the OPEN correction. a. Remove any device inserted in the test electrodes to create an OPEN condition (Nothing should be connected to the test electrodes). b. Press , and the following menu is displayed. c. Select OpenMeas using or , and press . After a while, the OPEN correction is completed. (If Out Of Limit is displayed, see \Performing the OPEN Correction |Canceling the stray admittance in parallel with the DUT" in Chapter 2.) 7. Perform the SHORT correction. a. Insert the shorting plate to the test xture as shown in the following gure: 3-8 Measurement Examples