Troubleshooting Service Function Front Panel Key Test Follow these steps to test the function of a front panel key. Step 1. Press the Config key (blue key, - key). Use the ® key to blink Svc and then press the Enter key. The service function menu below appears. Step 2. Use the ® key to blink KEY and then press the Enter key. The front panel key test screen appears. Step 3. Press a key you want to test. The name of the pressed key is displayed next to KEY TEST: in the upper line of the display. The following figure shows an example when the Trig key is pressed. If the name of the key is not displayed correctly, the key does not function. Contact your nearest Agilent Technologies sales office or the company from which you purchased this instrument. Step 4. If you want to finish the test, press the Enter key twice. The service function menu appears. Check that Exit is blinking (if not, operate keys to blink Exit), press the Enter key to return to the immediately previous menu, and then exit from this menu in the same way. Handler Interface Test Verify the Handler Interface Function using the handler interface simulator. For detail, refer to Handler Interface functional test in “Handler Interface Test” on page 37. Scanner Interface Test (Opt. 001 only) Verify the Scanner Interface Function using the scanner interface simulator. For detail, refer to Scanner Interface functional test in “Scanner Interface Test (opt. 001 only)” on page 41. 80 Chapter 4