Download Epson Powerspan Specifications

Transcript
Test Procedure (ZVM & ZVK)
5.2.2.2
ZVx
Linearity
Test equipment: 20 dB attenuator PC 3.5, SMA (ZVM) or PC 2.92 (ZVM, ZVK),
Test Cable ZV-Z14 (ZVM) or ZV-Z15 (ZVK),
Female-female adapter from ZV-Z32 (ZVM) or ZV-Z34 (ZVK)
Test setup:
Connect test cable with 20 dB attenuator between PORT1 and PORT2.
Settings on network analyzer:
PRESET
CAL
POWER UNCAL off
MODE
SWEEP MODE : POWER SWEEP
SWEEP
NUMBER OF POINTS = 26 (ZVM), 21 (ZVK)
SOURCE
STEP ATT A1 or STEP ATT A2 to 0 dB
SOURCE
Measurement frequency
START
-20 dBm
STOP
5 dBm (ZVM), 0 dBm (ZVK)
IF BANDWIDTH
10 Hz
MEAS
RATIO B1/A2 (Lin. PORT1)
MEAS
RATIO B2/A1 (Lin. PORT2)
FORMAT
MAGNITUDE, PHASE (Phase unwrap)
MARKER
MARKER CONT
MARKER1
= -10 dBm
DELTA MARKER
REF MARKER1
MARKER2
Measurement level
Reference
measurement:
½ Store trace and use it for normalization:
Magnitude: TRACE : DATA TO MEMORY : SHOW MATH (/)
Phase: TRACE : DATA TO MEMORY : SHOW MATH (-)
Measurement:
½
Remove attenuator, connect test cable directly to port.
½
Set reference value to 20 dB if required (better representation).
½
Record marker value at measurement level according to test record.
½
Repeat procedure for the other measurement frequencies.
1127.8500.60
5.10
E-2