Download Agilent Technologies B2902 Technical data
Transcript
Subsystem Commands :CALCulate:LIMit:COMPliance:DIGital[:DATA] :CALCulate[c]:FEED? Parameter type Data type. MATH|RESistance|CURRent|VOLTage (default). Parameter data type is CPD. type=VOLT specifies the voltage measurement data. type=CURR specifies the current measurement data. type=RES specifies the resistance calculation data given by the following formula. Resistance=Vmeas/Imeas Where, Vmeas is the voltage measurement data, and Imeas is the current measurement data. type=MATH specifies the data given by a math expression. The math expression must be specified before the :CALC:FEED MATH command is executed. An existing math expression can be specified by the :CALCulate:MATH[:EXPRession]:NAME command. A new math expression can be defined by the :CALCulate:MATH[:EXPRession]:NAME and :CALCulate:MATH[:EXPRession][:DEFine] commands. Query response type <newline> type returns the present setting of data type, MATH, RES, CURR, or VOLT. Response data type is CRD. Example :CALC:FEED MATH :CALC2:FEED? :CALCulate:LIMit:COMPliance:DIGital[:DATA] Defines the bit pattern used to indicate a failure of the compliance status check specified by m. It must be entered in the format set by the :FORMat:DIGital command. Syntax :CALCulate[c]:LIMit[m]:COMPliance:DIGital[:DATA] bit_pattern :CALCulate[c]:LIMit[m]:COMPliance:DIGital[:DATA]? Parameter bit_pattern 4- 10 0 (default setting in decimal expression) to 16383. Parameter data type is NR1 or NDN. Agilent B2900 SCPI Command Reference, Edition 2