Apply the gas to the instrument and regulate the cylinder output pressure so that the flow meter shows a
minimum flow (0,5 l/m) this ensures that the instrument is getting exactly the gas needed by its internal pump.
The instrument measures the concentration of the appllied gas; wait at least 3 minutes for the measure to
stabilize. The reading is shown in the 'Measured' row.
CO CALIBRATION
Action
calibrate
►Applied
1000.0 P
Measured
990.5 P
Is
82.22 uA
Ia
10.17 uA
Status
original
2
•
3
4
Whenever the reading is not satisfactory, proceed with the recalibration; Select the concentration of the applied
gas (cylinder concentration) in the 'Applied' row.
CO CALIBRATION
Action
calibrate
►Applied
Applied
1022.0 P
Measured
990.5 P
Is
82.22 uA
Ia
10.17 uA
Status
original
•
With
row.
keys set the cyclinder concentration in the 'Applied'
Select 'Calibrate' in 'Action' row.
CO CALIBRATION
►Action
Action
calibrate
Applied
1022.0 P
Measured
990.5 P
Is
82.22 uA
Ia
10.17 uA
Status
original
5
•
Depress the '
' key to save the new calibration.
CO CALIBRATION
6
►Action
calibrate
Applied
1022.0 P
Measured
990.5 P
Is
82.22 uA
Ia
10.17 uA
Status
cal ok
7
The recalibration result will be shown in the 'Status' row:
'CAL OK':
sensor recalibrated successfully
'CAL ERROR': sensor not recalibrated due to:
- The gas was not correctly delivered to the
instrument.
- The applied gas concentration was not correctly
set in the 'Applied' row.
- The 3 minutes stabilization time was not
observed.
- The sensor might be damaged or hexhausted
and need to be replaced.
It is always possible to bring back the original calibration with the action 'set original' and bring back the
last on-site calibration with 'set-user'.
8