Download Midmark M11 Specifications
Transcript
Maximum Capacities Load Type Packs (†) † M11/M11D Large / Deep Tray M11/M11D Small Tray M11/M11D Sterilizer Total 4425 cu. cm up to 5.0 cm thick (270 cu. in. up to 2 in. thick) 3195 cu. cm up to 5.0 cm thick (195 cu. in. up to 2 in. thick) 15240 cu. cm up to 5.0 cm thick (930 cu. in. up to 2 in. thick) Packs to have a minimum of 1/4 in. (6.4 mm) space between each other and away from all sterilizer surfaces. Operating the Sterilizer WARNING Do not use this sterilizer for sterilizing volatile substances or for any purpose other than its intended design. Burns and toxic or explosive conditions could result. Do not force door handle at any time. Chamber pressure may cause door to open with extreme force. If door handle does not move freely, allow unit to cool and depressurize for 40 minutes before opening door. Failure to adhere could result in serious personal injury. Operation CAUTION Programmable cycles 1 & 2 are provided for those applications requiring sterilization parameters different than the preset cycles. All material processed in these cycles must be validated by the user to ensure sterility of the processed load. 15