Download Agilent Technologies E4440
Transcript
One-Button Measurement Functions Meas Setup (Multi-Carrier Power—MCP) Example: MCP:CARR:AUTO 0 MCP:CARR:AUTO? MCP:CARR -100 MCP:CARR? 5.19.9 Limit Test Pressing Limit Test turns the testing of the limit line on or off. When Limit Test is set to On, each offset is compared to its upper and lower offset limit. In those cases where the power exceeds the limit, a red “F” is placed next to the dBc result to indicate a failure. If there are any failures, the pass/fail indicator area show “FAIL” in red; if there are none, it shows “PASS” in green. Any offsets that are in the off state (see “Offset Freq” on page 438) are not measured and their results will not be displayed on screen. Key Path: Meas Setup, More State Saved: Saved in instrument state. Factory Preset: Off History: Added with firmware revision A.02.00 One-Button Measurement Functions Chapter 5 443
Related documents
Agilent Technologies N8201A Technical data
U3700 Series User's Guide
Agilent Technologies E4406A VSA Satellite Radio User Manual
HP LaserJet 4250 User's Manual
User`s Guide
User`s/Programmer`s Reference Volume 1, Core
Agilent Technologies N9320A Technical data
CALCulate:DPSA:MARKer:PEAK:LEFT (No Query Form)
Agilent Technologies Model A.08.xx Water Dispenser User Manual
WCA230A & WCA280A Option 28 TD-SCDMA - TRS
Rockwell and Brinelll Hardness Testers
King Tester Portable Brinell Hardness Tester