Download 106-914-E Summit 11_12K User`s Guide
Transcript
12800 GUARDED THERMAL PROBE STATIONS The 12800 offers the same thermal testing capacity as a 12700, but with an advanced reduced-noise thermal chuck. Using Cascade Microtech’s patented FemtoGuard thermal chuck technology, the 12700 provides wafer temperature control in a reduced noise and capacitance environment. Chuck noise is reduced 1000-times over standard thermal chucks. The noise on topside probes reduces to fA levels. Chuck Specifications Flatness Residual capacitance, chuck to shield Capacitance variation over chuck surface Isolation, chuck to shield Breakdown bias voltage Chuck leakage current, thermal chuck on Chuck leakage current, thermal chuck off Temperature range Temperature uniformity Vacuum distribution area Auxiliary chucks 18 • Summit 11K/12K Probe Station User’s Guide 25-micron (1 mil) to 130°C, 51-micron (2 mil) to 200°C 1 pF 3 fF > 1 TΩ > 500-volts 0 to 100 volts: < 50 fA 0 to 100 volts, 20 fA -65 °C to 200 °C maximum (controller dependent). 300 oC on HT version. +0.5 °C or 0.5% whichever is higher 13, 75, or 152 mm (selectable) Two with individual vacuum control