Transcript
Troubleshooting Failure Message on Display TMS 1 The converter measures a heating current of less than 300 mA when heater is on. Possible Cause Measures Fuse TMS blown. Step 1: Check fuse Replace fuse TMS (F4A, 5x20 mm) if blown. Step 2: Check pump (heating element X31) Check if the connector of the heating element is connected. The resistance of the heating element X31 is typically between 70 Ω and 75 Ω. In case abnormal values are measured contact Leybold service. Step 3: Check TMS cable If step 2 was successful do the following: Check TMS cable for bent pins Measure resistance between pins X21/AA and X21/q with the cable connected to the pump. The resistance is typically between 70 Ω and 75 Ω. Replace the cable if it is damaged or in case the measurement of the resistance shows abnormal values now. Contact Leybold service. Heating element or internal pump connection damaged. TMS/DRIVE cable or connectors damaged. Converter failure. TMS 2 The converter measures a heating current of more than 4 A. Overvoltage. Heating element defective. TMS 3 TMS temperature sensor reads a temperature higher than 139 °C. TMS sensor damaged. Converter failure. Converter failure. TMS 4 TMS temperature sensor reads a temperature lower than 1 °C. TMS sensor short-circuited. Converter failure. Check the line voltage (200-240 V +10% / -15%). Repeat step 2of failure TMS 1. In case the measured value is less than 60 Ω contact Leybold service. Contact Leybold service. Check TMS sensor Measure the resistance between pins 1 and 3 of the TMS temperature sensor. The resistance is typically between 110 Ω and 150 Ω (20 °C to 140 °C). Contact Leybold if the sensor is defective. Contact Leybold service. Measure the resistance between pins 1 and 3 of the TMS temperature sensor. Contact Leybold if the resistance of the sensor is under 100 Ω. Contact Leybold service. GA05141_0702 - 09/2004 101