CM20 START-UP CHECK LIST
●
CCD camera control (left of microscope column):
Power on,
on PELTIER COOLER: COOL,
COOL
SHUTTER: AUTO, CAMERA: OUT.
●
On microscope column, Objective aperture and
Selected Area aperture out (levers to the right).
●
EDX detector is out of microscope
column: marker should point to
about 8.7 on the ruler. If EDX
detector in (marker points to much
lower number), press RETRACT
button on the Detector Position
Controller.
●
Manual tilt on sample goniometer is at 0° tilt; manual tilt
control of the goniometer is fixed – black lever pushed
towards the unit.
●
Check stage positions X: ~0.0 um, Y: ~ 0.0 um
If not, select XY CTRL to access XY control page; then press RESET
button, just below the CRT screen on the left, to zero X, Y.
Once reset complete, press READY button, just below CRT screen on the
right, to return to main CRT page.
1
2
3