Download User Guide
Transcript
Small Volume Direct Calibration Setup 1. Prepare the 9707BNWP combination nitrate electrode as described in the Electrode Preparation section and fill the electrode with Optimum Results F filling solution, Cat. No. 900046. 2. Connect the electrode to the meter. 3. Prepare at least two standards that bracket the expected sample range and differ in concentration by a factor of ten. Standards can be prepared in any concentration unit to suit the particular analysis requirement. See the Serial Dilution section for instructions on how to prepare standards. All standards should be at the same temperature as the samples. For details on temperature effects on electrode performance, refer to the Temperature Effects section. Nitrate Ion Selective Electrode User Guide 21