Download Quick Start User Manual
Transcript
Getting Acquainted with Your Instrument Timing Control You can set parameters for sampling rate or clock using the Timing page of Settings window. 1. Sets Sampling Rate. 2. You can set Repetition Rate when the Run mode is not set to Sequence. 3. You can select Clock Source (Internal or External). If External is selected, the clock signal from External Oscillator Input is used. If Internal is selected, the clock signal is generated internally. 4. Divider Rate can be set when the Clock Source is set to External. 5. You can select the Reference Source (Internal or External). Reference Source is selectable only when the Clock Source is set to Internal. 6. You can select the External Reference Type (Variable or Fixed). This parameter is selectable only when the Clock Source is set to Internal and the Reference Source is set to External. Quick Tips The Sampling Rate can be set when the internal clock source is selected and one of the following conditions is met: Internal is selected as the reference source. External is selected as the reference source and Fixed is selected as the external reference type. The arbitrary waveform generator accepts a 10 MHz, 20 MHz, or 100 MHz frequency signal as a Fixed external reference source. You can set the Multiplier Rate when the Clock Source is set to Internal, the Reference Source is set to External, and the External Reference Type is set to Variable. 34 AWG5000 and AWG7000 Series Quick Start User Manual