numbers). Figure 2-22 is an example of search results for a Tare Table,
with three views showing all available columns of data.
Figure 2-22: Tare Table Search Results
Calibration Test
The IND560 terminal provides a programmable Calibration Test sequence that can
contain up to 25 individual steps. The Calibration Test is designed to lead the
tester through a set of predetermined steps and compare the terminal calibration
with known test weights.
on the home
The test is accessed by pressing the CALIBRATION TEST softkey
page. A screen displays where the name of the tester can be recorded and the
serial number and capacity of each of the test weights used can be entered. This
provides a traceable record for the calibration test.
The calibration test is started by pressing the START softkey
. A display is
shown that provides the active scale weight, the target and tolerance weight
values, and two lines of instructions (as to where to place the weights). When the
tester completes these instructions, he or she presses the OK softkey
and the
terminal compares the actual weight on the scale to the programmed target weight
for this step. If the comparison passes, the test continues to the next step. If it fails,
2-32