Download User Guide - Wolf Laboratories

Transcript
Measuring a Sample Solution
8.4 Executing a Sample Measurement (with Printer, single-probe)
If your device is equipped with a printer (Option D) you can use it to log
the measurement results. The result of every sample will be printed if
you turn on the option USE PRINTER in the screen at the beginning of
each measurement run. Every sample shall get a unique identifier
(Sample ID). This can be done manually or automatically.
Fig. 12
1
measurement – printer
modus, single-probe
Before each sample series one should check the calibration
of the chloride meter by running a test measurement with a 100 mmol standard solution.
2
After successful measurement please press the SINGLE
SAMPLE button on the screen. If the automatic sample
numbering was activated you will be asked to inject the sample.
In manual sample numbering you will be asked to enter the
Sample ID and then to inject the sample. Alternatively you can
inject the sample directly without pressing any button. The
measurement then starts automatically and you will be asked for
Fig. 13
measurement –
probe numbering
Fig. 14
measurement –
sample injection
the Sample ID after the measurement has finished.
3
Now inject 20µl of the sample (sweat, urine, serum or rather a
standard or a control solution). Please consider the advices in
the chapter CONDITIONING respectively the correct way to
inject a sample.
4
The chloride meter automatically detects the injection and will start
the measurement. While the measurement is running the message CONDITIONING IS
RUNNING... will be shown on the display. You can track the measurement process by reading the
increasing measure value. If you like to abort the measurement please press the CANCEL button.
5
After the successful measurement the message
MEASUREMENT SUCCESSFUL and the measurement result
value in mmol/l will be displayed. Additionally the result will be
printed on paper.
6
If you like to measure another sample please just inject 20 µl of
the next sample into the measurement vessel. If you like to end
the measurement press the STANDBY button on the touch
screen to enter the standby mode and then press CANCEL.
Fig. 15
measurement –
sample reading
Measuring a Sample Solution | 21