Measuring a Sample Solution 8.4 Executing a Sample Measurement (with Printer, single-probe) If your device is equipped with a printer (Option D) you can use it to log the measurement results. The result of every sample will be printed if you turn on the option USE PRINTER in the screen at the beginning of each measurement run. Every sample shall get a unique identifier (Sample ID). This can be done manually or automatically. Fig. 12 1 measurement – printer modus, single-probe Before each sample series one should check the calibration of the chloride meter by running a test measurement with a 100 mmol standard solution. 2 After successful measurement please press the SINGLE SAMPLE button on the screen. If the automatic sample numbering was activated you will be asked to inject the sample. In manual sample numbering you will be asked to enter the Sample ID and then to inject the sample. Alternatively you can inject the sample directly without pressing any button. The measurement then starts automatically and you will be asked for Fig. 13 measurement – probe numbering Fig. 14 measurement – sample injection the Sample ID after the measurement has finished. 3 Now inject 20µl of the sample (sweat, urine, serum or rather a standard or a control solution). Please consider the advices in the chapter CONDITIONING respectively the correct way to inject a sample. 4 The chloride meter automatically detects the injection and will start the measurement. While the measurement is running the message CONDITIONING IS RUNNING... will be shown on the display. You can track the measurement process by reading the increasing measure value. If you like to abort the measurement please press the CANCEL button. 5 After the successful measurement the message MEASUREMENT SUCCESSFUL and the measurement result value in mmol/l will be displayed. Additionally the result will be printed on paper. 6 If you like to measure another sample please just inject 20 µl of the next sample into the measurement vessel. If you like to end the measurement press the STANDBY button on the touch screen to enter the standby mode and then press CANCEL. Fig. 15 measurement – sample reading Measuring a Sample Solution | 21