Download HMM100 User's Guide
Transcript
Chapter 6 ___________________________________________________________ Troubleshooting Problem or Message Cannot connect to serial line, current serial settings of the module are unknown. Likely Causes and Solutions Perform the following steps to connect: Any of the following errors is active: Internal error. Perform the following steps: 1. Set your terminal settings to 19200 8 N 1 (the default settings of the module). 2. Connect to the service port of the module. 2. Power up the module, and type "z" on the serial line at least five times. 4. The module will come online with the default settings, and with the text SERVICE MODE. 5. Use the SERI command to set the desired serial settings, and use the SAVE command to save them. 1. Reset or power cycle the module. Check if the - Program checksum error error is gone. - Parameter storage 2. Return the module to factory settings using the checksum error FRESTORE command. Check again. - Sensor coefficient checksum 3. If the error is still active, contact a Vaisala error service center. - Current settings checksum error - Default settings checksum error - Log corrupted - Resistance measurement Technical Support For technical questions, contact the Vaisala technical support via email: helpdesk@vaisala.com For contact information of Vaisala Service Centers, see http://www.vaisala.com/services/servicecenters.html VAISALA _______________________________________________________________________ 63