Download BD FACSAria User's Guide
Transcript
Setting Up for Sorting Into a Plate or Slide This section describes how to set up for sorting into a plate or slide. For general guidelines, see Setting Up for Sorting on page 138. Installing the Sorting Hardware B WARNING Any instrument surface that comes in contact with biological specimens can transmit potentially fatal disease. Use universal precautions when handling sorting hardware. Wear suitable protective clothing and gloves. 1 Install the splash shield below the aspirator drawer. Close the sort block door and open the sort collection chamber door, if needed. The sort block door must be closed in order to open the collection chamber door. Remove the tube holder, if one is installed, and tilt the splash shield to insert it into the slotted fittings below the sort aspirator drawer. Push the splash shield all the way in. splash shield Figure 5-3 Installing the splash shield Chapter 5: Sorting 153