Setting Up for Sorting Into a Plate or Slide
This section describes how to set up for sorting into a plate or slide. For general
guidelines, see Setting Up for Sorting on page 138.
Installing the Sorting Hardware
B WARNING
Any instrument surface that comes in contact with biological
specimens can transmit potentially fatal disease. Use universal precautions when
handling sorting hardware. Wear suitable protective clothing and gloves.
1 Install the splash shield below the aspirator drawer.
Close the sort block door and open the sort collection chamber door, if
needed. The sort block door must be closed in order to open the collection
chamber door.
Remove the tube holder, if one is installed, and tilt the splash shield to
insert it into the slotted fittings below the sort aspirator drawer. Push the
splash shield all the way in.
splash shield
Figure 5-3 Installing the splash shield
Chapter 5: Sorting
153