Download BD FACSAria User`s Guide
Transcript
Optimizing the Signal from the Violet Laser Repeat the steps in Optimizing Signal from the Red Laser on page 123, using the Violet worksheet and Violet tube. Tracking QC Results To track instrument performance over time, copy PMT voltages, means, and CVs for each fluorescence parameter onto a QC log such as the one provided on page 147. Alternatively, print the experiment’s instrument settings or Instrument Status report and a copy of each worksheet and keep a copy of the printouts. The instrument settings printout contains PMT voltages and compensation settings; the Instrument Status report includes the sheath pressure and flow rate, lasers used, laser delay, area scaling, and window extension settings in addition to PMT voltages and compensation settings. • To print instrument settings, right-click the Instr Settings icon for the experiment and choose Print, or select the icon and click Print in the Inspector. • To print the report, set the current tube pointer on the last-used tube and choose Instrument > Instrument Status Report. Choose File > Print Report in the report window. Reusing the Experiment To reuse this experiment for QC on another day, do the following. 1 Open the instrument QC experiment. 2 Right-click the most recent specimen and choose Duplicate without Data. The three QC tubes appear under the new specimen. 3 Rename the specimen with the current date. Chapter 4: Running Samples 125