Description Resistance Measurements four when the 25/second reading rate is selected. Resolution versus range and reading rate are shown in Table 1. Table 1. Reading rate for various ranges Range Resolution 6/s Rate Resolution 25/s Rate 200 mV 10 uV 50 uV 2V 100 uV 500 uV 20 V 1 mV 5 mV 200 V 10 mV 50muV 500 V 100 mV 500 mV Resistance Measurements The resistance mode operates by forcing a calibrated value of current through the unknown resistance and measuring the consequent voltage drop. The ohmmeter is autoranging and will select either the 200 W, 2 kW, 20 kW, 200 kW, or 2 MW range depending on the value of the unknown resistance being measured. It may also be fixed on any range. Reading rates of 6 readings per second or 25 readings per second may be selected. Software features also permit displaying a computed value which is a function of resistance, using user-entered values of offset and scale parameters. The DCX-127 input configuration permits making these resistance measurements on either a 2-wire or 4-wire basis. DCX-127 Multifunction Module User’s Guide Page 3