Appendix B GSIOC Control GSIOC Commands GSIOC commands Non-specific commands The 306 software includes a set of commands that offers direct compatibility with first generation Gilson pumps. Command Type ? B D F L S U Z or $ I B B B B B B I Function Request Status Begin Dispense Cycle (s) Set Dispense Mode Set Flow Mode Lock Pump Stop Flow Unlock Pump Master Reset All commands are detailed below with their type, mode and function. Immediate commands have their response format described. Buffered command parameters are documented. Comments are added where necessary. Immediate ? Request Status Response format: abcdefgh where ‘a’ is the error status ‘L’ for low pressure limit ‘H’ for high pressure limit ‘I’ for Invalid settings ‘ ’ (blank) for no error. ‘b’ is the control status ‘L’ for ‘locked’ and ‘U’ for ‘unlocked’. ‘cdefg’ being the current parameters and ‘h’ the mode status ‘F’ for Flow mode ‘M’ for Microflow mode ‘D’ for Dispense mode ‘S’ for Stop (not in dispense mode) Buffered B Begin Dispense Cycle (s) Syntax: Bn Parameter: n is the number of dispense cycles. If n is omitted, it is considered as 1 by default. Buffered D Set Dispense Mode Syntax: D Comment: this command sets the pump in Dispense mode ; it does not initiate liquid flow. Appendix B-3