71 Placement and calibration of sensors
Calibrating Photo-Ionization Detector (PID) Sensor
Calibrating Photo-Ionization Detector (PID) Sensor
PID parts per million (ppm) zero calibration
It is preferred to flush the PID ppm sensor with Zero Grade Air for a zero
calibration. Alternatively, you may flush the PID ppm sensor with
Nitrogen. In clean environments (with no presence of toxic chemicals,
such as in a school setting with no toxic cleaners present) applying room
air can be used to set your zero calibration point.
For the flow rate, it is recommended to set it at 1 LPM for best response
time.
PID parts per billion (ppb) zero calibration

NOTE for PID ppb sensor: The PID ppb sensor may be susceptible to
baseline settlement after more than 2 weeks of storage. In these cases,
the PID PPB sensor should be operated in clean air for several hours prior
to performing a calibration and prior to performing measurements of low
concentration of gases (< 1000 PPB).
The PID ppb sensor uses a static base line for the zero offset. If the user
adjusts zero base line, the PID ppb needs to be enabled in the setup
menu.
 To enable, select Setup from the main screen and then choose PID. In
the PID screen, enable the ppb zero feature. Once set, return to the
calibration screen to resume the calibration process. (For more details
on setting, please refer to “PID sensor setup” in Chapter 3.)
 PID ppm/ppb zero cal
1. Connect the Zero Grade Air and cal cup to the EVM. (Please skip this
step if you are not applying a gas.)
2. From the start screen, press the Down
Calibration and press Enter
.
3. Press the
arrow to select
Arrows to select the PID sensor.
Selecting PID sensor
Cal softkey
Figure 4-27: Selecting PID sensor for calibration process