Download Simultaneous User Manual
Transcript
71 Placement and calibration of sensors Calibrating Photo-Ionization Detector (PID) Sensor Calibrating Photo-Ionization Detector (PID) Sensor PID parts per million (ppm) zero calibration It is preferred to flush the PID ppm sensor with Zero Grade Air for a zero calibration. Alternatively, you may flush the PID ppm sensor with Nitrogen. In clean environments (with no presence of toxic chemicals, such as in a school setting with no toxic cleaners present) applying room air can be used to set your zero calibration point. For the flow rate, it is recommended to set it at 1 LPM for best response time. PID parts per billion (ppb) zero calibration NOTE for PID ppb sensor: The PID ppb sensor may be susceptible to baseline settlement after more than 2 weeks of storage. In these cases, the PID PPB sensor should be operated in clean air for several hours prior to performing a calibration and prior to performing measurements of low concentration of gases (< 1000 PPB). The PID ppb sensor uses a static base line for the zero offset. If the user adjusts zero base line, the PID ppb needs to be enabled in the setup menu. To enable, select Setup from the main screen and then choose PID. In the PID screen, enable the ppb zero feature. Once set, return to the calibration screen to resume the calibration process. (For more details on setting, please refer to “PID sensor setup” in Chapter 3.) PID ppm/ppb zero cal 1. Connect the Zero Grade Air and cal cup to the EVM. (Please skip this step if you are not applying a gas.) 2. From the start screen, press the Down Calibration and press Enter . 3. Press the arrow to select Arrows to select the PID sensor. Selecting PID sensor Cal softkey Figure 4-27: Selecting PID sensor for calibration process