Download Simultaneous User Manual

Transcript
APPENDIX
A
126 Appendix A
Specifications
Particulate Sensor
Method
Photometer
Particulates
Display
Range
0.000- 200.0 mg /m3
(Non-condensing)
0 to 20,000 ug/m3
0.1 – 10 μm
Display
Resolution
0.001
1
N/A
Accuracy/
Repeatability
+/-15% (Calibrated to Arizona
road dust; ISO 12103-1, A2 Fine
Test Dust)
*See note below
Particulate Sensor - General
Size
0.1 μm to 10 μm* (*Note: the photometer can detect particles up
to 100 μm; however, accuracy is reduced for sizes greater than 10
μm It may not be as accurate.)
Impactor- mechanical
Top-mounted, rotary impactor particulate filter
Impactor settings
PM2.5, PM4, PM10 or PM at 1.67 Liter/minute flow rate
Measures
Real-time aerosol/dust concentration using a 90º optical light
scattering photometer to report total mass concentration (mg/m3 or
ug/m3) of particulate matter
Gravimetric sampling


Dust correction factors
Life expectancy/Warranty
Flow: ± 5%, after calibration
Back-pressure @ gravimetric cassette
 ≥ 5 kPa
 ≥ 20 inches of water @ 1.67 lpm
Up to 8 User-defined particulate correction factors. To correct for
variation from “Arizona Road Dust”.
5,000 hours (for the laser photo-meter)/1 year warranty