Download Simultaneous User Manual
Transcript
APPENDIX A 126 Appendix A Specifications Particulate Sensor Method Photometer Particulates Display Range 0.000- 200.0 mg /m3 (Non-condensing) 0 to 20,000 ug/m3 0.1 – 10 μm Display Resolution 0.001 1 N/A Accuracy/ Repeatability +/-15% (Calibrated to Arizona road dust; ISO 12103-1, A2 Fine Test Dust) *See note below Particulate Sensor - General Size 0.1 μm to 10 μm* (*Note: the photometer can detect particles up to 100 μm; however, accuracy is reduced for sizes greater than 10 μm It may not be as accurate.) Impactor- mechanical Top-mounted, rotary impactor particulate filter Impactor settings PM2.5, PM4, PM10 or PM at 1.67 Liter/minute flow rate Measures Real-time aerosol/dust concentration using a 90º optical light scattering photometer to report total mass concentration (mg/m3 or ug/m3) of particulate matter Gravimetric sampling Dust correction factors Life expectancy/Warranty Flow: ± 5%, after calibration Back-pressure @ gravimetric cassette ≥ 5 kPa ≥ 20 inches of water @ 1.67 lpm Up to 8 User-defined particulate correction factors. To correct for variation from “Arizona Road Dust”. 5,000 hours (for the laser photo-meter)/1 year warranty