Download Levelflex M (Operating Instructions)

Transcript
Levelflex M FMP 40 with Foundation Fieldbus
6.7
6 Commissioning
Function "envelope curve display" (0E3)
You can obtain the following information from the envelope curve display in this function:
full calibr.
quality of
evaluated echo
empty calibr.
evaluated echo
is marked
envelope curve
minimum distance
of the plot
full calibr.
distance of
evaluated echo
probe length
empty calibr.
maximum distance
of the plot
level echo
mapping
empty calibr.
level echo
substracted signal
probe length
Navigation in the envelope curve display
Using navigation, the envelope curve can be scaled horizontally and vertically and
shifted to the left or the right. The active navigation mode is indicated by a symbol in the
top left hand corner of the display.
Horizontal Zoom mode:
- horizontal zoom in
- horizontal zoom out
Move mode:
- moved to the left
- moved to the right
Vertical Zoom mode:
…
Endress+Hauser
- vertical zoom (4 steps)
69