Download 100432-an-01-ml-Scha..
Transcript
When the display appears in full, first release the MAXHLD button and then the ON/OFF button Calibreur acoustique The automatic switch-off is decommissioned, the device remains switched on until the ON/OFF button is pressed. The switch-off function is automatically reactivated when the device is next switched on. Handling Never operate the unit immediately after bringing it from a cold into a warm area. The resulting condensation water could damage the unit. Allow the unit to come to room temperature before switching it on. Wait until condensation evaporates. Avoid operation under unfavourable ambient conditions. These will result in damage to the sensitive electronics inside the measuring device or sensor. Unfavourable ambient conditions include: • excessive humidity (> 90 % relative humidity), • moisture • dust and combustible gases, fumes or solvents • excessive ambient temperatures (> +50°) • excessively low ambient temperatures (< 0 °C) Never expose yourself to high sound levels without officially approved, tested hearing protection. Otherwise, your hearing may be damaged. Replace the device in the carrying case for storage. This will protect the device from damage. Vue de dessous Tournevis de réglage RS232 interface The device has an interface which works according to the RS232 standard. The sound level meter can be connected to a computer via this interface. The determined data can be stored in the computer. The relevant connection cable and software are available. 56 41