TECHNICAL INFORMATION
The following example command is used to read the
current of the channel 1 or 2.
Ascii Protocol:
@ 04 T02 ? 0409
40h 30h 34h 54h 30h 32h 3Fh 30h 34h 30h 39h
Binary Protocol:
$ 10 T0 2 4.4 E-04 0679
24h 31h 30h 54h 30h 32h 34h 2Eh 34h 45h 2Dh 30h
34h 30h 36h 37h 39h
ü 04 T02 ? 1 |
81h 30h 34h 54h 30h 32h 3Fh 6Ch
• 10 T02 8.9E – 04
F
Multigauge Compatible Protocol:
1h 31h 30h 54h 30h 32h 38h 2Eh 39h 45h 2Dh
30h 34h 15h
#108?
23h 31h 30h 38h 3Fh Dh
> 1 08 1.9 E - 04
3Eh 31h 30h 38h 31h 2Eh 39h 45h 2Dh 30h 34h Dh
High Voltage Commands
Command
Description
Mode
Channels
Format
Possible values
fixed/step
fixed/step mode
Read
Write
HV1, HV2
Status
30h fixed
31h step
start/protect
start/protect mode
Read
Write
HV1, HV2
Status
30h start
31h protect
Psel
Power supply polarity
Read
HV1, HV2
Status
30h negative
31h positive
Vmax
Vmax variable
Read
Write
HV1, HV2
Integer (V)
[3000, 7000] step 100
Imax
Imax variable
Read
Write
HV1, HV2
Integer (mA)
[100, 400] step 10
Pmax
Pmax variable
Read
Write
HV1, HV2
Integer (W)
[100, 400] step 10
Iprotect
Iprotect variable
Read
Write
HV1, HV2
Integer (mA)
[10, 100] step 10
Vstep1
Vstep1 variable
Read
Write
HV1, HV2
Integer (V)
[3000, 7000] step 100
Istep1
Istep1 variable
Read
Write
HV1, HV2
Exp. (A)
[1.0E-9, 1.0E1]
Vstep2
Vstep2
Read
Write
HV1, HV2
Integer (V)
[3000, 7000] step 100
Istep2
Istep1 variable
Read
Write
HV1, HV2
Exp. (A)
[1.0E-9, 1.0E1]
SetPt1
SetPt1 variable
(Set Point 1)
Read
Write
HV1, HV2
Exp. (Torr)
[1.0E-9, 1.0E1]
(have to be greater than
SP2)
SetPt2
SetPt2 variable
(Set Point 2)
Read
Write
HV1, HV2
Exp. (Torr)
[1.0E-9, 1.0E1]
Remote I/O
Output
Reads the status of the Read
Remote I/O outputs
HV1, HV2
BitField
See Remote I/O table
Remote I/O
Input
Reads the status of the Read
Remote I/O inputs
HV1, HV2
BitField
See Remote I/O table
63/73