SC1100 Calibration Option
SC1100 Calibration Setup
7
Table 7-1. SC1100 Calibration and Verification Equipment (cont.)
Leveled Sine Wave Frequency, Time Marker Verification
Instrument
Model
Minimum Use Specifications
Frequency Counter
PM 6680 with option
(PM 9621, PM 9624, or
PM 9625) and (PM
9690 or PM 9691)
2 ns to 5 s, 50 kHz to 600 MHz: <0.15 ppm
uncertainty
Adapter
Pomona #3288
BNC(f) to Type N(m)
SC1100 Cable (N-BNC)
(supplied with SC1100)
Type N to BNC
Wave Generator Verification
AC Measurement
Standard
Fluke 5790A with -03
option
Range
1.8 mV p-p to 55 V p-p
Frequency
10 Hz to 100 kHz
Adapter
Pomona #1269
BNC(f) to Double Banana Plug
Feedthrough 50 Ω ±1 %
Termination
SC1100 Cable (N-BNC)
(supplied with SC1100)
Type N to BNC
SC1100 Calibration Setup
The procedures in this manual were made to let users calibrate the SC1100 at their own
site if it becomes necessary to do so. It is strongly recommended that, if possible, you
send your Calibrator to Fluke for calibration and verification. The Calibrator Mainframe
must be fully calibrated before you do calibration of the SC1100.
The hardware adjustments are intended to be one-time adjustments done in the factory.
Some times adjustment can be necessary after repair. Hardware adjustments must be
done before calibration. Calibration must be done after if hardware adjustments are made.
See the “Hardware Adjustments” section in this chapter.
The AC Voltage function is dependent on the DC Voltage function. Calibration of the
AC Voltage function is necessary after the DC Voltage is calibrated.
The Calibrator Mainframe must complete a warm-up period and the SC1100 must be
turned on for a minimum of 5 minutes before you start calibration. This lets internal
components become thermally stable. The Calibrator Mainframe warm-up period is a
minimum of two times the period the calibrator was turned off, or a maximum of 30
minutes. Push c to turn on the SC1100. The green LED on the SCOPE key is illuminated
when the SC1100 is turned on.
Most of the SC1100 Option can be calibrated from the front panel. Push  to turn on
the SC1100 and wait a minimum of 5 minutes. To start the Scope Cal mode:
1. Push .
2. Push the CAL softkey.
3. Push the CAL softkey again.
4. Push the SCOPE CAL softkey.
Note
If you push the Scope Cal softkey sooner than 5 minutes after you pushed
, a warning message shows in the display.
7-15