Download T/PUMP®

Transcript
SERVICE MANUAL
TP600 & 700 Series T/PUMP
DISASSEMBLY/REASSEMBLY
Heater Assembly:
•
Perform the Steps for Reservoir and Upper Front & Rear Housing
disassembly.
• Remove the 2 wires from the Cartridge Heater. One goes to the
PC-Board, the other goes to a neutral wiring connector.
Remove the (2) brass fittings on the bottom of the Tray assembly which
holds the brass manifold block to the tray.
To assembly, reverse the disassembly steps. Special attention is required
for the following:
1. Gray hose can be placed in warm water (<50°C) to facilitate
installation on manifold fittings.
2. All o-rings must be seated. The reservoir o-ring is shaped, and shape
must be aligned with the reservoir.
3. All fittings on the manifold should have teflon tape applied to prevent
leaks.
4. Insure wires are cleared from between enclosures and away from
enclosure screws.
5. Insure wires are routed away from the cooling fan blades.
6. For proper screw size and location for reservoir and housing, refer to
parts illustration and replacement parts list.
www.gaymar.com
25