System FiVe Software Upgrade to v.1.9
e GE Vingmed Ultrasound
c.) Write down the System Software Name listed in the Module Software
Revisions List
File name/ date
System FiVe :
(System/options)
(
)
Options :
Table 0-1 System Software Name listed in the Module Software Revisions List.
Note:
Membership and Advantage will need a password for all functions covered
by the introducing offering on Advantage:
Functions:
•
•
•
•
•
•
7.) [Sw.1.7.x only] Select the Applications menu when having the 2.5 FPA probe
connected, The following software options must be verified and noted for:
- RF,
- Contrast,
- TVI,
- or any combination of these options.
Section 8.0 - Install Memory Upgrade
[Sw.1.7.x only]
1.) Turn of the power (rear of system).
2.) Remove both the Upper and the Lower Covers on the left side.
3.) Un-screw the retaining screws used for fastening the System Processor
Board and pull out the board.
4.) Un-screw the screws holding the RAM piggyback and remove the piggyback.
5.) Gently install the new memory piggy back and fasten it with the screws that
were used for the old piggy back.
6.) Insert the Processor board in the board rack, in the same position it was
before. Be sure that it is seated properly. Fasten the retaining screws.
7.) Turn on the power and boot the scanner.
8.) Verify that all the boards starts (the red LEDs are turned off) and that the
system boots completely.
9.) Turn off the system and mount the panels you removed earlier.
System FiVe - Service Manual - FA091050 rev. I
N-27