9 • Troubleshooting
Recognizing
module
Error
code
B
19
B
Art.-Nr.: 80116-271
01.07
Dok.-Nr: D010050-EN, Ver.: 007, ÄM-Nr: 09088, Gültig ab: 26.02.07, Gedruckt: CDIERL/15.09.08, Ausdruck nicht maßstäblich und kein Original.
Recognizing module: B = Control panel
1A
Additional information
Number of the incorrect parameter (Hex):
0: Activation assignment
3: CutMode
4: CutIntensity
5: CutEffect
6: CoagIntensity
7: CoagEffect
8: CoagMode
B: CutAPCFlow
C: CoagAPCFlow
D: CutIESFlow
E: CoagIESFlow
F: IES-Basic Suction
10: IES-Run-On Time
Description
Remedy
PowerFail:
Occurs if a PowerFail signal
is received but the PowerFail does not actually happen within 2 s.
Check house installation.
Error in parameter memory:
Occurs if an error is identified by the checksum
parameter monitoring of a
program. (e.g. a bit flips in
DRAM memory).
One-off event -> EMC problem => Check environment.
Check low-voltage power
supply unit.
Frequent event -> Error in
system memory =>
Replace assembly.
Reproducible -> Software
error => Inform ERBE
Tübingen Technical Service.
B
1B
CRC check not yet completed: Occurs if the user
wants to operate the unit
after downloading software
before there has been at
least one successful CRC
check.
Wait until the CRC check
has been completed.
B
1C
The maximum ON time has
been exceeded: Occurs if
activation lasts longer than
the ON time selected in
SET-UP.
User error.
B
1D
Incorrect instrument parameters:
Occurs if an instrument is
detected via instrument
recognition and the parameter is outside the permissible tolerances.
Check instrument software
version. Replace instrument. Check IIF.
Number of the incorrect parameter (Hex):
101 Cut-Mode
102 Coag-Mode
201 Cut-Effect
202 Coag-Effect
301 Cut-Power
302 Coag-Power
501 Cut-APC-Flow
502 Coag-APC-Flow
67 / 144