Download material specification for document 205522
Transcript
Continuum MR Infusion System and Wireless Remote Accessory Periodic Maintenance The following maintenance should be done on an annual basis: WARNING: Electrical Shock hazard: Operator injury could result from cleaning or servicing the system while powered. DO NOT immerse system components into any type of liquid. In order to avoid shock and prevent damage to the system, always disconnect from line power and remove batteries before cleaning or servicing. CAUTION: Any adjustments, maintenance, or repair of the pump should be performed by MEDRAD trained representatives only. Any adjustments, maintenance, or repair of the pump by a non-MEDRAD trained representative may impair the operation of the Continuum MR Infusion System and/or the accuracy of an infusion. 1. Replace the pump battery as explained in Installing or Replacing a Pump Battery. 2. Calibrate the pump as explained in Calibrating a Pump. MEDRAD recommends returning the pump to MEDRAD for general preventative maintenance on an annual basis (usually in conjunction with calibration). The pump is specified for a continuous use life of three years and should be replaced through MEDRAD service at the end of this time. 4-3