7. Remove the 2x screws securing the closing plate, and pull the plate forward from
the top, this gains access to the overheat switch, the lamp fixing and the PT100.
NOTE: The overheat switch is designed to cut the power to the lamp if the control
circuit fails and the software can no longer control the heater, it is designed to
operate at chamber temperatures of around 240 deg C. If the switch was to fail
OPEN, then the lamp would not light.
Other causes of lamp failure could be the lamp itself being faulty or either of the
micro switches either failing or being incorrectly adjusted. Any of these faults would
need the rework method described.
8. To remove the PT100, slide out the cable securing clamp from the PT100 side if the
hinge, the PT100 should now pull out from the wire location.
9. Close the lid carefully, and turn the lid assembly upside down.