Download CONTENTS
Transcript
OPERATING INSTRUCTIONS Page: 21 FUNCTION AND RANGE HINTS AND COMMENTS When active, trigger levels will be displayed according to this table: Use Read-Level to check what trigger-levels Auto-Level has selected. DC AC Auto Level Level selected by Auto. Level selected by auto. Manual Level set using Set Value. 0 V. When active, the signal on Input-A is also connected to Input-B. You can use COM B via A and the manual trigger level settings to make accurate rise-time measurements: The DC/AC and Attenuator switches for Input-A affects both 1. Input-A and B. The Input-B DC/AC and Attenuator switches have no effect. (See figure 7). 2. Trigger level/sensitivity and Slope can be selected separately. 3. 4. 5. 6. Select AC-coupling, COM B via A and positive slope for input A and B. Select VOLT A MAX-MIN and read the amplitude of the signal. Calculate 10 % of the peak to peak voltage. Depress SET A, use and to set trigger level A to the MIN reading plus the 10 % of Vpp. Depress SET B, use and to set trigger level B to the MAX reading minus the 10 % of Vpp. Select TIME A-B. This is the HF-input which must be used when the FREQ-C RATIO C/A and RATIO C/B can be selected via the bus. function is selected. If the Timer/Counter does not include the Input-C option, the BNC-connector is replaced by a plastic plug. Range: 70 to 1300 MHz. Impedance: 50Ω Sensitivity: 10 mVRMS up to 900 MHz, 15 mVRMS 900-1100 MHz and 40 mVRMS above. Max voltage: 12 VRMS The Timer/Counter automatically detects if a suitable signal is connected to the EXT- REF Input-connector. Use external reference when the measurement requires ultra-high stability. Suitable signal: 10 ± 0.1 MHz, 0.5 to 15 Vrms Sine wave. The Timer/Counter must still have the internal time base even if an external reference frequency is used. If single is selected, the EXT REF indicator on the display is not switched on until after the first measurement. PM 6666 - OPERATORS MANUAL