Download PhysioTrak
Transcript
MR3 Filter Mode The MR3 Filter Mode provides the best ECG performance during Cardiovascular MR procedures that involve steady state free precession imaging with balanced gradient (True- FISP, FIESTA, Balanced FFE) sequences. MR4 Filter Mode The MR4 Filter Mode provides the best ECG performance on 3.0 Tesla MR systems. PACER PULSE This selection is not available in the MDCU. ST DISPLAY This selection is not available in the MDCU. CAL Selecting this menu option sends a 1 mm/mV Calibration Pulse to the Recorder and to the ECG Waveform on the display screen. This feature may be used for analysis of the patient’s ECG. RETURN Selecting this menu option will return the monitor to the Normal Screen. 5.1.3 Patient Preparation Before placing the electrodes, the skin should be properly prepared. WA R N I N G Proper patient preparation is essential for prevention of skinburns. Low resistance of the electrodes must be guaranteed. The instructions below must be followed carefully. Prepare the patient as follows: Remove chest hair by shaving, if applicable. Clean the skin with special abrasive skin prepping gel (e.g. Nuprep from D.O. Weaver and Company) and a gauze pad. The skin may turn slightly red because of rubbing. Use a clean gauze pad to dry the area thoroughly. P HYSIO TRAK Release 1.0 5-7 5