Download here - DennLec

Transcript
Model 8642A/B
Performance Tests
Measurement
1. Re-connect the instruments as shown in Figure 4-8 Setup A.
2. Set the signal generator and H P 8642 frequency to the value listed in the table below.
3. Set the signal generator level to the Sig Gen Level recorded in the table below.
4. Set the H P 8642 level to the values given in the table below.
5. On the H P 8642 enable the phase increment mode (SHIFT SPCL 250) and use the knob to vary
the phase. As the phase of the H P 8642 is varied the readings on the measuring receiver should
change.
6. Record the maximum and minimum readings on the measuring receiver display in the table below
(Measurement Readings).
7. Subtract the Maximum from the Minimum reading in and record the Result in the table below
(Results M a x - M i n ) .
Frequency
(MHz)
Hi
8. Verify that the actual result is less than the specified Maximum.
Cal. 1
Reading
Cal. 2
Reading
Measurement
Readings
Sig
Gen
Level
Min
Max
Results
Max-Min
Actual
Maximum
4.000
4.000
4.000
4.000
10.0
6.0
-1.0
-6.0
6.02
6.02
3.52
3.52
dB
dB
dB
dB
1000.000
1000.000
1000.000
1000.000
10.0
6.0
-1.0
-6.0
6.02
6.02
3.52
3.52
dB
dB
dB
dB
2000.000
2000.000
2000.000
2000.000
10.0
6.0
-1.0
-6.0
6.02
6.02
3.52
3.52
dB
dB
dB
dB
NOTE
If actual SWR values are needed, compute SWR using the following formula:
SWR
= \C\(((rnax~rn'i'n
*n
dBm))/2Q)
Low Level SWR
1. Connect the instruments as shown in Figure 4-8 Setup B.
2. Preset the instruments.
3. Set the Signal Generator to:
+16 dBm
.. 4 MHz
LEVEL
FREQUENCY
rev.08JUL88
Performance Test 8
4-29