Download File

Transcript
Data Sheet
Characteristics
Characteristic
General
Characteristic
Description
Number of
Channels
TLA5201B
TLA5202B
TLA5203B
TLA5204B
Time Stamp
Clocking/Acquisition
Modes
(all channels are acquired including clocks)
34 channels (2 are clock channels)
68 channels (4 are clock channels)
102 channels (4 are clock and 2 are qualifier channels)
136 channels (4 are clock and 4 are qualifier channels)
51 bits at 125 ps resolution (3.25 days duration)
Asynchronous and Synchronous. 125 ps (8 GHz) MagniVu™
high-speed timing is available simultaneous with all modes
Input Characteristics (with P64xx probes)
Characteristic
Description
Capacitive Loading
(P6419, P6450)
(P6410, P6434)
Threshold Selection
Range
<0.7 pF typical data/clock
2 pF typical data/clock
From –2.0 V to +4.5 V in 5 mV increments.
Threshold presets include TTL (1.5 V), CMOS (1.65 V), ECL
(–1.3 V), PECL (3.7 V), LVPECL (2.0 V), LVCMOS 1.5 V
(0.75 V), LVCMOS 1.8 V (0.9 V), LVCMOS 2.5 V (1.25 V),
LVCMOS 3.3 V (1.65 V), LVDS (0 V), and user defined
Threshold Selection Separate selection for each of the clock/qualifier channels
Channel Granularity and one per group of 16 data channels
Threshold Accuracy ±(100 mV)
(including probe)
Input Voltage Range
–2.5 V to 5.0 V
Operating
Nondestructive
±15 V
Minimum Input Signal Swing
P6410, P6419,
±250 mV
P6450
P6434
±300 mV
Input Signal
200 mV/ns typical
Minimum Slew Rate
2
State Acquisition Characteristics
www.tektronix.com
Description
235 MHz
Maximum State
Clock Rate
Maximum State
470 Mb/s
Data Rate
4/2 Mb , 16/8 Mb, 64/32 Mb
State Record
Length with Time
Stamps (half/full
channels)
16 ns range that may be shifted towards the setup region by
Setup-and-Hold
0 ns [+8, -8] ns, 4 ns [+12, -4] ns, or 8 ns [+16, 0] ns
Time Selection
Range
Setup-and-Hold
All Channels: 1.5 ns typical
Window
Minimum Clock Pulse Width
P6434
1.5 ns
P6410, P6419,
1.25 ns
P6450
Demux Channel
Channels can be demultiplexed to other channels through
Selection
user interface with 8-channel granularity
Timing Acquisition Characteristics
Characteristic
Description
MagniVu™ Timing 125 ps (8 GHz).
Resolution
Storage rate adjustable to 250 ps, 500 ps, 1 ns, and 2 ns
MagniVu Timing
16 Kb per channel, with adjustable trigger position
Record Length
Timing Resolution 500 ps/1 ns/2 ns to 50 ms
(quarter/half/full
channels)
Timing Record
8/4/2 Mb, 32/16/8 Mb, 128/64/32 Mb per channel
Length
(quarter/half/full
channels with time
stamps and with or
without transitional
storage)
Timing Record
Half of default main record length
Length with Glitch
Storage Enabled
Channel-to-channel 1 ns (900 ps typical)
Skew
Minimum Recognizable Pulse/Glitch Width (single channel)
P6410, P6419,
1 ns
P6450
P6434
1.25 ns
250 ps
Minimum
Detectable
Setup/Hold Violation
Minimum
Sample period + channel-to-channel skew
Recognizable
Multichannel Trigger
Event