Download File
Transcript
Data Sheet Characteristics Characteristic General Characteristic Description Number of Channels TLA5201B TLA5202B TLA5203B TLA5204B Time Stamp Clocking/Acquisition Modes (all channels are acquired including clocks) 34 channels (2 are clock channels) 68 channels (4 are clock channels) 102 channels (4 are clock and 2 are qualifier channels) 136 channels (4 are clock and 4 are qualifier channels) 51 bits at 125 ps resolution (3.25 days duration) Asynchronous and Synchronous. 125 ps (8 GHz) MagniVu™ high-speed timing is available simultaneous with all modes Input Characteristics (with P64xx probes) Characteristic Description Capacitive Loading (P6419, P6450) (P6410, P6434) Threshold Selection Range <0.7 pF typical data/clock 2 pF typical data/clock From –2.0 V to +4.5 V in 5 mV increments. Threshold presets include TTL (1.5 V), CMOS (1.65 V), ECL (–1.3 V), PECL (3.7 V), LVPECL (2.0 V), LVCMOS 1.5 V (0.75 V), LVCMOS 1.8 V (0.9 V), LVCMOS 2.5 V (1.25 V), LVCMOS 3.3 V (1.65 V), LVDS (0 V), and user defined Threshold Selection Separate selection for each of the clock/qualifier channels Channel Granularity and one per group of 16 data channels Threshold Accuracy ±(100 mV) (including probe) Input Voltage Range –2.5 V to 5.0 V Operating Nondestructive ±15 V Minimum Input Signal Swing P6410, P6419, ±250 mV P6450 P6434 ±300 mV Input Signal 200 mV/ns typical Minimum Slew Rate 2 State Acquisition Characteristics www.tektronix.com Description 235 MHz Maximum State Clock Rate Maximum State 470 Mb/s Data Rate 4/2 Mb , 16/8 Mb, 64/32 Mb State Record Length with Time Stamps (half/full channels) 16 ns range that may be shifted towards the setup region by Setup-and-Hold 0 ns [+8, -8] ns, 4 ns [+12, -4] ns, or 8 ns [+16, 0] ns Time Selection Range Setup-and-Hold All Channels: 1.5 ns typical Window Minimum Clock Pulse Width P6434 1.5 ns P6410, P6419, 1.25 ns P6450 Demux Channel Channels can be demultiplexed to other channels through Selection user interface with 8-channel granularity Timing Acquisition Characteristics Characteristic Description MagniVu™ Timing 125 ps (8 GHz). Resolution Storage rate adjustable to 250 ps, 500 ps, 1 ns, and 2 ns MagniVu Timing 16 Kb per channel, with adjustable trigger position Record Length Timing Resolution 500 ps/1 ns/2 ns to 50 ms (quarter/half/full channels) Timing Record 8/4/2 Mb, 32/16/8 Mb, 128/64/32 Mb per channel Length (quarter/half/full channels with time stamps and with or without transitional storage) Timing Record Half of default main record length Length with Glitch Storage Enabled Channel-to-channel 1 ns (900 ps typical) Skew Minimum Recognizable Pulse/Glitch Width (single channel) P6410, P6419, 1 ns P6450 P6434 1.25 ns 250 ps Minimum Detectable Setup/Hold Violation Minimum Sample period + channel-to-channel skew Recognizable Multichannel Trigger Event