EEG troubleshooting
 To isolate a p
problem in the diff amp
p of p
particular channel and
its corresponding electrode, the calibration switch can be
used.
 If the
th channel
h
l under
d test
t t can be
b calibrated
lib t d with
ith the
th switch,
it h
this would indicate that the electronics beyond it is working
and that the problem would probably be in a lead.
 The frequency response of filters can be checked by
measuring the output of the filter due to a variable frequency
input.
input
 The most vulnerable element in active filters is the diff-amp
chip.
p Failure in the chip
p would cause a radical change
g in the
frequency response. However, if the changes in frequency
response are small, the problem may be a leaky capacitor, or
aging or thermal damage in the components.