Cranex D
5. Adjusting the alignment of the unit
7. Turn the rotating unit 180°. Make sure that the
black marks, previously placed on the pulley and
pulley holder are in the same place. This ensures that the rotating unit has not moved in the
backwards / forwards direction.
8. Press the exposure switch and check where the
beam strikes. The distance between the first
position, marked with the slider, and the second
position shows how much the rotating center
needs to be adjusted.
9. To adjust the position of the rotation center
loosen the three fixing screws (A) (one shown in
the picture) that hold the collimator guide in
position and turn the set screw (B) in the required direction.
Tighten the fixing screws and then check the
beam position again (steps 6 - 8). Readjust if
necessary.
10. Now adjust the beam so that it strikes the O line.
Push the rotating unit towards the column and
then press the RETURN button to drive the unit
to the PIO position.
11. On top of the rotating unit, loosen the two nuts
(B) that hold the Y-sensor in place. Turn screw
(A) to adjust the position of the interrupt plate
which will change the position where the rotating
unit stops in the PIO position, and thus change
the position of the beam.
Repeat steps 10 and 11 until the beam coincides with the O line.
Installation and set-up manual 8201085
43