FSP
Analyzer - MEAS
Power Measurements – Hardkey MEAS
With its power measurement functions the FSP is able to measure all the necessary parameters with
high accuracy in a wide dynamic range.
A modulated carrier is almost always used (except e.g. SSB-AM) for high-frequency transmission of
information. Due to the information modulated upon the carrier, the latter covers a spectrum which is
defined by the modulation, the transmission data rate and the signal filtering. Within a transmission
band each carrier is assigned a channel taking into account these parameters. In order to ensure errorfree transmission, each transmitter must be conforming to the specified parameters. These include
among others:
•
the output power,
•
the occupied bandwidth, i.e. the bandwidth which must contain a defined percentage of the power
and
•
the power dissipation allowed in the adjacent channels.
Additionally the menu contains functions to determine the modulation depth of AM modulated signals
rd
and to measure the 3 order intercept point.
The measurements and the corresponding settings are selected in the MEAS menu.
MEAS menu:
BW
SWEEP
MEAS
TRIG
TIME DOM
POWER
CHAN PWR
ACP
The MEAS key opens the menu to select and set the power
measurement.
The following measurements can be selected:
•
Power in the time domain (TIME DOM POWER)
•
Channel power and adjacent-channel power in the
frequency
domain
with
a
single
carrier
(CHAN PWR ACP)
•
Channel power and adjacent-channel power in the
frequency domain with several carriers
(MULT CARR ACP)
•
Occupied bandwidth (OCCUPIED BANDWIDTH)
•
Carrier-to-noise ratio (C/N, C/No)
•
Amplitude probability distribution (SIGNAL STATISTICS)
•
Modulation depth (MODULATION DEPTH)
•
3 order intercept (TOI)
OCCUPIED
BANDWIDTH
SIGNAL
STATISTIC
C/N
C/No
MODULATION
DEPTH
rd
TOI
SELECT
MARKER
1164.4556.12
The above measurements are carried out alternatively.
4.15-1
E-1