AUTOMATIC POWER DOWN The Automatic Power Down feature of the PRC 2000 system is a safety feature which removes power from all channels 90 minutes after all Active Channels have entered the Temperature Setback Mode. This feature is not programmable and is automatically enabled when all Active Channels are in the Temperature Setback Mode. OPERATION When all Active Channels have entered Automatic Temperature Setback Mode, a 90 minute timer within the system circuitry will start running. 1. If the system has no Active Channels (no connected handpieces), Automatic Shutdown will not occur. 2. If any key is pressed during the 90 minute time period, the timer is reset. Timer will not start timing out again until all Active Channels are in Temperature Setback Mode. 3. During the last minute before the 90 minutes expire, an audible tone will occur every 4 seconds to alert the operator. 4. At the end of the 90 minute period, the system will enter the Power Down Mode. Power is removed from all channels, all Channel LEDs will stop flashing and the Digital Readout will display a flashing “OFF”. EXITING POWER DOWN MODE Power Down can be exited with all channels returning to normal operation by pressing any key or by turning the POWER Switch OFF (“0”) and then back ON (“1”). 72