Download Manual - HMS Electronics

Transcript
Shimadzu Medical Systems
National Technical Support
20101 S. Vermount Ave. Torrance, Ca. 90502
Tel. (800) 228-1429
Fax. (310) 217-0729
SERVICE BULLETIN
Model: HMS Starter and UD150L-F/RII
Date: 10/19/98
Number: XR-980015
Title: HMS starter interface to the UD150L-F/RII
Checked by:__________
Approved by:__________
Issued by: Gordon Bowller
======================================================================
UD150L-F or UD150L-RII
HS2 Interface PCB
5V
JS1
J2
Install Jumpers for 15V
XUD CONT-94 PCB
U1
6
R1
2
4
1
R3
1K
C1
33
2
5.6K
D1
R2
B8
ST
Normal Speed
Rotation Start
2
5
R4
5V
1K
1N4004
510
*= no connections
with
UD150L-RII
3
U2
6
R5
2
4
1
R7
1K
C2
33
4
5.6K
D2
R6
FL
**=no connections
for a one X-ray
tube systems
5
R8
5V
1K
1N4004
510
U3
6
R9
2
4
1
R11
1K
C3
33
6
5.6K
D3
R10
5
R12
5V
1K
1N4004
510
A9
High Speed
Rotation Start
H2
HS
A1 B1 +15
7
+15V DC
U4
6
R13
2
4
1
C4
33
R15
1K
8
5.6K
D4
R14
5
R16
5V
5
1K
1N4004
510
x1 XTV TERMINAL-94 PCB
SF
F
9
U5
6
R17
2
4
1
R19
1K
C5
33
L0
10
5.6K
D5
R18
5V
6
1K
1N4004
510
*
TC2
5
R20
*
11
SP1
*
U6
R21
2
4
1
C6
R23
1K
33
D6
Tube 2 selected
B7
2M
19
18
A6
7
Normal Speed
Confirmation
20
B6
H7
High Speed
Confirmation
21
A15 B15
22
A5 B5
HON
1
A3 B3
+15V DC
12
5.6K
R22
TC1
**
5
R24
1K
1N4004
510
+24V DC
13
15
Kaux
INTLK
GND
0V
CAUTION: DO NOT MIX +15V DC AND L100 POWER SUPPLIES
VERIFY ALL JUMPERS ON THE HMS INTERFACE PCB
Reference the connection diagrams in the UD150L-F/RII schematics
for XUD CONT-94 (7/13) and XTV TERMINAL-94 PCB's.
1
page