Download FSP40
Transcript
Specifications Specifications are valid under the following conditions: 15 minutes warm-up time at ambient temperature, specified environmental conditions met, calibration cycle adhered to, and total calibration performed. Data without tolerances: typical values only. Data designated "nominal" applies to design parameters and is not tested. Data designated "σ = xx dB" is shown as standard deviation. ¸FSP3 ¸FSP7 ¸FSP13 ¸FSP30 ¸FSP40 9 kHz to 3 GHz 9 kHz to 7 GHz 9 kHz to 13.6 GHz 9 kHz to 30 GHz 9 kHz to 40 GHz Frequency Frequency range Frequency resolution 0.01 Hz Internal reference frequency (nominal) Aging per year1) 1 × 10 –6 Temperature drift 1 × 10–6 With option ¸FSP-B4 (OCXO) Aging per year1) 1 × 10–7 Temperature drift 1 × 10–8 External reference frequency 10 MHz Frequency display with marker or frequency counter Marker resolution Max. deviation (sweep time >3 × auto sweep time) span/500 ±(frequency × reference frequency + 0.5% × span + 10% × resolution bandwidth + ½ (last digit)) Frequency counter resolution 0.1 Hz to 10 kHz (selectable) Count accuracy (S/N >25 dB) ±(frequency × reference frequency + ½ (last digit)) Frequency span Max. span deviation 0 Hz, 10 Hz to 3 GHz 0 Hz, 10 Hz to 7 GHz 0 Hz, 10 Hz to 13.6 GHz 0.1% Spectral purity (dBc (1 Hz)) SSB phase noise, f = 500 MHz, for f > 500 MHz see diagrams below Carrier offset 100 Hz <–84, typ.–90 1 kHz <−100, typ.−108 10 kHz <−106, typ. −113 100 kHz2) <−110, typ. −113 1 MHz2) <−120, typ.−125 10 MHz typ.−145 Residual FM f = 500 MHz, RBW 1 kHz, sweep time 100 ms 1) 2) 2 After 30 days of operation. Valid for span >100 kHz. ¸FSP Spectrum Analyzer typ. 3 Hz 0 Hz, 10 Hz to 30 GHz 0 Hz, 10 Hz to 40 GHz