How to Manually Tune the Control Loop WARNING Do not place any part of your body in the path of a moving actuator. A crush zone exists between the actuator and any equipment in the path of its movement. Immediate and unexpected actuator response is possible when you apply hydraulic pressure to your system. Stay clear of the actuators when applying hydraulic pressure. 4. Turn on hydraulic pressure (see page 54). 5. Select the force control mode. Standalone FlexTest SE Controllers: Select Tuning > Manual > Control Mode > Force. Automated Controllers: A. In the Station Controls panel toolbar, click to display the Manual Command window. B. In the Manual Command window, select the appropriate Channel associated with the strain sensor signal you are tuning. C. For the Control Mode, select Force. 6. Install a dummy or broken specimen. Standalone FlexTest SE Controllers: A. Enable the Dial. B. Adjust the Dial to position the actuator during specimen installation. Automated Controllers: 84 Tuning Procedures A. On the Manual Command window. Click Enable Manual Command. B. Adjust the Manual Command window slider to position the actuator during specimen installation. MTS Series 793 Tuning and Calibration