Download - OMEGA Engineering
Transcript
Omega Model CYD211 User’s Manual Table 1-2. Sensor Input Performance Chart Sensor Type Silicon Diode GaAlAs Diode Temperature Coefficient Sensor Units Input Range Negative Volts (V) 0 – 2.5 V 10 µA ±0.01% 100 µV DT-670-SD with 1.4H Cal. 1.4 – 475 K DT-670 Negative Volts (V) 0 – 7.5 V 10 µA ±0.01% 100 µV TG-120SD with 1.4H Cal. 1.4 – 475 K Requires Calibration –31.6 mV at 4.2 K –1.73 mV at 77 K –2.3 mV at 300 K –2.12 mV at 500 K –180 mV/K at 10 K –1.25 mV/K at 77 K –2.75 mV/K at 300 K –2.75 mV/K at 475 K 20 µV 0.6 mK at 4.2 K 11.6 mK at 77 K 8.7 mK at 300 K 9.4 mK at 500 K 20 µV 1 mK at 10 K 16 mK at 77 K 10 mK at 300 K 10 mK at 475 K ±160 µV ±0.01% RDG ±10 mK at 4.2 K ±152 mK at 77 K ±94 mK at 300 K ±80 mK at 500 K ±160 µV ±0.02% RDG ±6 mK at 10 K ±300 mK at 77 K ±150 mK at 300 K ±110 mK at 475 K ±10 µV ±5 PPM of reading per °C ±31 mK at 4.2 K ±267 mK at 77 K ±154 mK at 300 K ±140 mK at 500 K ±20 µV ±5 PPM of reading per °C ±21 mK at 10 K ±390 mK at 77 K ±140 mK at 300 K ±210 mK at 475 K Recommended for T > 60 K & B < 3 T Recommended for T > 4.2 K & B < 5 T Sensor Excitation (Constant Current) Display Resolution (Sensor Units) Example LSCI Sensor Temperature Range Standard Sensor Curve Typical Sensor Sensitivity Measurement Resolution: Sensor Units Temperature Equivalence Electronic Accuracy: Sensor Units Temperature Equivalence Temperature Coefficient Temperature Accuracy including electronic accuracy, CalCurve™ and calibrated sensor Magnetic Field Use 1-6 Introduction