Download ConST 116 - Insatech
Transcript
Basic Operation Connection A Open to atmosphere DIGITAL PRESSURE C ALIBRATOR B DIGITAL PRESSURE C ALIBRATOR Units Units SW SW V mA COM V mA COM 24V 24V ConST273 ConST273 open pressure switch dial gauge Pre-pressure pressure transmitter C flexible hose Zeroing should be operated under this condition only. D Fine adjust process open Pressurizing process E R DIGITAL PRESSURE CALIBRATO DIGITAL PRESSURE CALIBRATOR R DIGITAL PRESSURE CALIBRATO Units Units Units operate it to get the ⑤ desired value (0~6)MPa (0~6)MPa (0~6)MPa operate it to get the ② desired value ① close,do not use too much force ④ close,do not use too much force ③ open set it to the mid ② position of its travel ① Open ② fine adjust ① push the security bolt pin in Decrease pressure process F G Vacuum process Repeat Vacuum process DIGITAL PRESSURE CALIBRATOR H R DIGITAL PRESSURE CALIBRATO DIGITAL PRESSURE CALIBRATOR Units 0.06 Units Units 0.04 0.06 真空表 (0~6)MPa 0.04 真空表 08 0.02 - 0.1 0 (0~6)MPa 08 0.02 - 0.1 0 (0~6)MPa it slowly to get the desired ① open value, and close! ② fine adjust pull out ② operate it to get the ⑤ desired value operate it to get the ② desired value it must be open when ④ close(but select the pressure/vacuum) ① open ① open ③ push in,select the vacuum position Decrease vacuum process I Calibration procedure 1st point R DIGITAL PRESSURE CALIBRATO Units 0.06 Pressure 0.04 真空表 (0~6)MPa 08 0.02 - 0.1 0 open it slowly to get the desired ② value, and close! ① close ③ fine adjust N point : A → B → C → D → E → D … E → D → F Dial gauge Pressure transmitter 1st point Vacuum 2nd point 2nd point N point : A → B → G → D → H → D … H → D → I Pressure : A → B → C → D Vacuum : A → B → G → D Pressure switch increase pressure till the switch move increase pressure till the switch move F → D F → D decrease pressure till the switch resume decrease pressure till the switch resume the end the end Remark: ConST group has made a concerted effort to provide complete and current information for the proper use of the equipment. The product specifications and other information contained this manual are subject to change without notice. Beijing ConST Instruments Technology Inc.