Download ConST 116 - Insatech

Transcript
Basic Operation
Connection
A
Open to atmosphere
DIGITAL PRESSURE C
ALIBRATOR
B
DIGITAL PRESSURE C
ALIBRATOR
Units
Units
SW
SW
V
mA
COM
V
mA
COM
24V
24V
ConST273
ConST273
open
pressure switch
dial gauge
Pre-pressure
pressure transmitter
C
flexible hose
Zeroing should be operated under
this condition only.
D
Fine adjust process
open
Pressurizing process
E
R
DIGITAL PRESSURE CALIBRATO
DIGITAL PRESSURE CALIBRATOR
R
DIGITAL PRESSURE CALIBRATO
Units
Units
Units
operate it to get the
⑤ desired
value
(0~6)MPa
(0~6)MPa
(0~6)MPa
operate it to get the
② desired
value
① close,do not use too much force
④ close,do not use too much force
③ open
set it to the mid
② position
of its travel
① Open
② fine adjust
①
push the security bolt pin in
Decrease pressure process
F
G
Vacuum process
Repeat Vacuum process
DIGITAL PRESSURE CALIBRATOR
H
R
DIGITAL PRESSURE CALIBRATO
DIGITAL PRESSURE CALIBRATOR
Units
0.06
Units
Units
0.04
0.06
真空表
(0~6)MPa
0.04
真空表
08
0.02
- 0.1
0
(0~6)MPa
08
0.02
- 0.1
0
(0~6)MPa
it slowly to get the desired
① open
value, and close!
② fine adjust
pull out
②
operate it to get the
⑤ desired value
operate it to get the
② desired
value
it must be open when
④ close(but
select the pressure/vacuum)
① open
① open
③
push in,select the vacuum position
Decrease vacuum process
I
Calibration procedure
1st point
R
DIGITAL PRESSURE CALIBRATO
Units
0.06
Pressure
0.04
真空表
(0~6)MPa
08
0.02
- 0.1
0
open it slowly to get the desired
② value,
and close!
① close
③ fine adjust
N point
: A → B → C → D → E → D … E → D → F
Dial gauge
Pressure transmitter
1st point
Vacuum
2nd point
2nd point
N point
: A → B → G → D → H → D … H → D → I
Pressure
: A → B → C → D
Vacuum
: A → B → G → D
Pressure switch
increase pressure
till the switch move
increase pressure
till the switch move
F → D
F → D
decrease pressure
till the switch resume
decrease pressure
till the switch resume
the end
the end
Remark: ConST group has made a concerted effort to provide complete and current information for the proper use of the equipment. The product specifications
and other information contained this manual are subject to change without notice.
Beijing ConST Instruments Technology Inc.