Download EC-Lab User`s Manual

Transcript
EC-Lab Software User's Manual
Disconnect the cables from the cell, select Electrode
connections = -3V,+14 V control and reconnect the cell
as follows:
- P1, S1 and S2 leads to the working electrode
- S3 and P2 leads to the counter electrode
P.S: The impedance techniques are not available with 3V;+14V control mode.
Fig. 58:-3V;+14V control mode
for SP-300 technology.
2.6 Accepting and saving settings and running a technique
2.6.1 Accepting and saving settings
Once you have set the parameters of your experiment, click on the Accept icon
. It validates the Parameters Settings, the Cell Characteristics and the Advanced Settings and
sends them to the instrument. If the option is activated in the Config\option\Warning menu,
the following window will appear before accepting the settings:
Fig. 59: Summary window before accepting settings (VMP3 technology).
This window summarizes several parameters of the experiment. Click on Yes to accept the
settings and start the experiment. The settings can be set as default settings for the current
technique. Right-click on the mouse and select “Set settings as Default”. The parameter settings can be saved as an *.mps file in Experiment\save as\ or right-click on Save Experiment…, or click on
.
2.6.2 Running an experiment
Click on the Run button
40
. A filename selection window will appear: