Download EC-Lab Techniques and App Manual
Transcript
Techniques and Applications Manual Fig. 55: DNPV waveform. Fig. 54: DNPV detailed diagram. The scan increment is defined by a pseudo staircase made of steps of amplitude PH and duration ST. As mentioned above only one point is recorded at the end of the potential forward pulse and one point at the end of the potential reverse pulse, making two points during the ST period. The settings above (Fig. 54) are given for a positive scan. To perform a negative scan set Ev inferior to Ei and SH to a negative value. average I over the last ... % of each step ( ... points) Selects the end part of the potential step for the current average (<I>) calculation, to exclude the first points where the current may be disturbed by the step establishment. A value of 100 % will take all the step points for the average and a value of 0 % will take only the last point. Note that the current average (<I>) is recorded at the end of the potential step into the data file. Scan rate = … mV/s … number of points ~ These values are given as an indication and are calculated by the computer. The scan rate is directly given by SH / (0.001ST) and the number of points is roughly 2(Ev - Ei)/ST for the forward scan. E Range = … enables the user to select the potential range and to adjust the potential resolution according to the experiment (See EC-Lab Software User’s Manual for more details on the potential resolution adjustment). I Range … Bandwidth … sets the current range and bandwidth values for the entire experiment. Note: It is highly recommended to avoid using the automatic current range with pulsed techniques. The resolution of each range is different and dynamic current range changes may lead to spikes on the plot. 65