Download Triggered - Astronics Test Systems
Transcript
Publication No. 980897 Rev. A Test Results 3156B User Manual Pass Fail 2. Remove the cable from 3156B channel 1 and connect to chan 2 3. Repeat the test procedure as above for channel 2 Test Results Pass Fail Equipment: Oscilloscope AM Preparation: 1. Configure the oscilloscope as follows: Time Base: 0.5 ms Trigger source: Channel 2, positive slope Amplitude: 1V/div 2. Connect 3156B Channel 1 output to the oscilloscope input, channel 1 3. Connect the 3156B SYNC output to the oscilloscope input, channel 2 4. Configure the 3156B channels 1/2 as follows: Reset Wave Mode: Modulation Modulation Mode: AM Carrier Frequency: 1MHz Modulating Frequency: 1kHz Modulation Depth: 500kHz Sync: On Output: On Enable: On Test Procedure 1. Verify AM operation on the oscilloscope as follows: Waveform: Amplitude Modulated Sine Modulation Depth: 10kHz Test Results Pass Fail 2. Remove the cable from 3156B channel 1 and connect to chan 2 3. Repeat the test procedure as above for channel 2 Test Results Astronics Test Systems Pass Fail Maintenance and Performance Checks 6-29