Download pdf
Transcript
Filters Typical life span of a prefilter is 3 months but differing room conditions can require replacement more or less frequently. During the first 3 months of operation the pre-filter should be checked once per month to determine the average life span for your facility. Regular certification will dictate life spans and replacement schedules of the supply HEPA filter. Regular replacement of the prefilter will lengthen the life of your supply HEPA filter. One digital pressure gauge is an indicator of differential pressure between the blower inlet and the Supply HEPA filter. It is a qualitative indicator of the HEPA filter condition. Your certifier should note the starting pressure and monitor it periodically. A large change in readings in a short time period is indicative of possible failures. Typical HEPA filter life span is 3-5 years. The second digital gauge measures the differential pressure between the airlock and the work area. Two-Part Glove System We recommend a glove with a longer cuff length for use with the VERSAFLOW. The sleeve and “bracelet” system will accommodate almost any commercially available glove. Typical life span of the sleeves is up to 6 months and GERMFREE recommends that they are changed at this interval. A visual inspection of all glove and sleeve surfaces is recommended at the beginning and end of each day. Any tears or punctures would require a replacement. With the unit turned on the operator can run a hand on all surfaces to feel for any air moving into or out of the work area through imperfections in the surfaces. Soap solution can also be used. If the VERSAFLOW has pressure the soap will bubble at any cut, tear or puncture. 18