Download document - Voss Scientific

Transcript
correct for the components depends on both the signal type (e.g., wideband waveform, narrowband waveform, image) and the component type.
For example, before a raw waveform is acquired by an instrument, suppose it passes
through a sensor, an attenuator and a 10-foot cable. The raw data acquired by the instrument is
not the same as was seen at the sensor because each component has an effect on the original signal. CalMan reduces the data by using the original captured input signal then compensating for
the loss through the cable and attenuator. The sensor reduction is based on the specific sensor
type used, and would likely involve a conversion of units (e.g. to power density, E-field, amps).
CalMan precisely determines the data conversion by applying each item’s reduction algorithm,
which is used in conjunction with the item’s characterization data.
Item Data pages for many Cal Items include a curve specification that characterizes the
item’s effect on the input signal. Selecting Curve in the item interface activates the curve dialog,
which either creates a polynomial or selects a calibration waveform file. Preferably, this curve
should be the waveform that resulted from a direct characterization of the particular device, such
as a network analyzer sweep.
When performing data reduction on Cal Items that do not represent physical components,
CalMan is simply applying user-specified functions in an automated fashion. Common user
processing includes waveform scaling, integration, dividing one channel by another, creating
overlays and extracting rows from images. There are three primary advantages to CalMan’s
method of data processing: 1) it automates the entire data processing effort – once CalMan is
configured, multiple data channels over multiple acquisitions can be processed with a few mouse
clicks; 2) it provides complete documentation – a permanent record of the processing steps is
associated with the processed data; and 3) it uses a simple graphical interface that does not require the user to write code.
6.1.4
Direction and Input / Output naming conventions
Nodes may have a notion of signal inputs and outputs (or also parent and child), but it is
important to understand that the input and output sides may change depending on the context.
For example, examine the CalMan graph shown in Figure 6-1 with digitizers on the left side and
sensors on the right. During data acquisition data flows from right to left (from the sensor to the
digitizer), so the node outputs are on the left side. During data reduction data flows from left to
right (digitizer to sensor), so outputs are on the right side. A physical component like a splitter
has multiple outputs that would be drawn on the left side. A processing component like an image branch has multiple outputs on the right side. Also note that some cal items, such as
switches, can be oriented with their multiple ports on either side. Physical inputs and outputs are
referred to as ports, which indicates flow from sensor to digitizer. Data processing node inputs
and outputs will be preceded by the tag “reduction”, indicating flow from digitizer to sensor.
The following section presents particular Cal Items or classes of Cal Item that have multiple input and/or output ports.
6.1.4.1
Multi Input / Output nodes
The majority of the cal items have a single input and a single output, whether referring to
their physical connections or data flow during reduction. Here the special cases are presented –
those nodes that have either multiple inputs or multiple outputs (at present no items are supported that have both).
DAAAC 4.0 User’s Manual
112