Download Solomat - INLEC.com
Transcript
General Description Table 1 .1 Environmental measurements using the MPM4100 Measurement parameter Relative humidity Dewpoint temperature NOTE : Recommended probe (Section 1 .5) Measurement range 355RH, 356RH, 357RH, 358RH 0 to 100% RH 226RH 10% to 70% RH As above -35°C to +70°C A Pt100 temperature sensor within the RH probe measures air temperature, which is used to calculate dewpoint temperature Temperature 200 series, type-K probe -80°C to +900°C 100 series probes (Pt100) (mechanically configured for particular applications) -190°C to +600°C Hotwire airspeed probe 127MS or 129MS/GN 0 .1 m/s to 12 m/s (0°C to 50°C) Vane airspeed probe 228MS or 228GN 0.5 m/s to 40 m/s (-10°C to +70°C) Pitot static tube 50-4, 380, 229MS, 229ML, 229MM 1 .5 m/s to 100 m/s NOTE: You must use a 511 LP pressure probe in conjunction with pitot static tubes Volume flow rate (vfr) Pitot static tube, hotwire or vane Determined by the crosssectional area of the duct Temp/humidity/wind speed Solomat 3-element thermal probe, type 1010TC -10°C to +70°C, 0 to 100%RH and 0.1 to 12 m/s Tachometer speed 327TM, 427TM 6 rpm to 20,000 rpm Frequency Any frequency or event-time source 0 .5 Hz to 1 MHz (see Section 1 .3.3) Air pressure 511 LP 512HP 515BP ±30 in . H20 0 to 150 PSIG 0 to 1 .15 bar absolute Water pressure 513LW 514HW ±150 in . H20 0 to 300 PSIG C02 concentration 1201 GS 0 to 5000 ppm CO concentration 1212GS 0 to 500 ppm Customer's own probe See Section 1 .3 .3 Iss 2 (Apr 94) Page 1 -7