Download USER MANUAL FOR MICROPROCESSOR

Transcript
Cycle parameters within each program are as follows.
(a)
STERILIZING TEMPERATURE. This is the drain temperature which
is also used as the control temperature during this phase of the cycle.
The available range of temperatures is 100 C to 140°C.
(b)
PURGE METHOD. There are two available methods for purging either
Vacuum or Steam purge.
(c)
VACUUM PURGE. If vacuum purge is selected the number of purge
pulses may be set. The range of available pulses is 1 to 20 .
OPTION. If the optional pressure transducer has been fitted you will
be able to set the purge pulse depth. This consists of two values for
purge pulse low and purge pulse high. The displayed values are
absolute quantities.
(d)
STEAM PURGE. During the air removal stage free steaming
(purging) is undertaken for a fixed time. This parameter is the required
purge time, and values from 1 to 30 minutes can be set. NOTE. Due to
the high pressure steam generator fitted to this machine, purge times
tend to be closer to those required by a direct steam machine rather
than an internal steam generator.
(e)
HOLDING TIME. The period of time for which the sterilising
temperature is maintained. The holding time will start when both drain
and load temperatures are equal to or greater than the sterilising
temperature. The maximum available holding time is factory set
according to the type of machine.
(f)
SAFE LOAD TEMPERATURE. This parameter allows a choice of
cycle finishing temperatures. It is sensed from the load simulator. This
parameter has an available range of 50 to 90°C.
(g)
LOAD FAIL TEMPERATURE. The user may program in the
minimum temperature at which, during the holding time, a faulty cycle
is declared, should the load temperature fall below this value. The
default value is sterilising temperature minus 2 C. The range is from
the sterilising temperature to ten degrees below the sterilising
temperature.
(h)
START RECORDING TEMPERATURE. The printer will start to
record the drain and vessel temperatures at the set level. The default
value will be sterilising temperature minus 2°C, but any value from
20°C to set point minus 2°C may be set.
(i)
START/STOP RECORDING TEMPERATURE. If your autoclave is
fitted with the optional extra of a temperature chart recorder, the chart
will start and stop at the same values of temperature as are programmed
into the start and stop printing temperature parameters. If this option is
fitted continuous recording will be made on the chart recorder, and
21