Cycle parameters within each program are as follows. (a) STERILIZING TEMPERATURE. This is the drain temperature which is also used as the control temperature during this phase of the cycle. The available range of temperatures is 100 C to 140°C. (b) PURGE METHOD. There are two available methods for purging either Vacuum or Steam purge. (c) VACUUM PURGE. If vacuum purge is selected the number of purge pulses may be set. The range of available pulses is 1 to 20 . OPTION. If the optional pressure transducer has been fitted you will be able to set the purge pulse depth. This consists of two values for purge pulse low and purge pulse high. The displayed values are absolute quantities. (d) STEAM PURGE. During the air removal stage free steaming (purging) is undertaken for a fixed time. This parameter is the required purge time, and values from 1 to 30 minutes can be set. NOTE. Due to the high pressure steam generator fitted to this machine, purge times tend to be closer to those required by a direct steam machine rather than an internal steam generator. (e) HOLDING TIME. The period of time for which the sterilising temperature is maintained. The holding time will start when both drain and load temperatures are equal to or greater than the sterilising temperature. The maximum available holding time is factory set according to the type of machine. (f) SAFE LOAD TEMPERATURE. This parameter allows a choice of cycle finishing temperatures. It is sensed from the load simulator. This parameter has an available range of 50 to 90°C. (g) LOAD FAIL TEMPERATURE. The user may program in the minimum temperature at which, during the holding time, a faulty cycle is declared, should the load temperature fall below this value. The default value is sterilising temperature minus 2 C. The range is from the sterilising temperature to ten degrees below the sterilising temperature. (h) START RECORDING TEMPERATURE. The printer will start to record the drain and vessel temperatures at the set level. The default value will be sterilising temperature minus 2°C, but any value from 20°C to set point minus 2°C may be set. (i) START/STOP RECORDING TEMPERATURE. If your autoclave is fitted with the optional extra of a temperature chart recorder, the chart will start and stop at the same values of temperature as are programmed into the start and stop printing temperature parameters. If this option is fitted continuous recording will be made on the chart recorder, and 21